您的位置:山东大学 -> 科技期刊社 -> 《山东大学学报(医学版)》

山东大学学报(医学版)

• 论文 • 上一篇    下一篇

Pentacam与IOLMaster测量角膜曲率及前房深度的比较

潘虹,王利华,马鲁新   

  1. 山东大学附属省立医院眼科中心, 济南 250021
  • 收稿日期:2008-01-19 修回日期:1900-01-01 出版日期:2008-06-16 发布日期:2008-06-16
  • 通讯作者: 王利华

Pentacam and IOLMaster in keratometry and anterior chamber depth measurements: a contrast study

PAN Hong, WANG Li-hua, MA Lu-xin   

  1. Ophthalmic Center, Provincial Hospital Affiliated to Shandong University, Jinan 250021, China
  • Received:2008-01-19 Revised:1900-01-01 Online:2008-06-16 Published:2008-06-16

摘要: 目的比较Pentacam三维眼前节分析诊断系统与IOLMaster光学相干生物测量仪测量角膜曲率及前房深度的差异,并对两种测量结果进行一致性评价。方法受检者年龄(20.80±1.29)岁,等效球镜屈光度(-3.44±2.04)D。应用Pentacam和IOLMaster分别对双眼角膜曲率和前房深度进行测量,均取受试者的左眼测量数据进行统计,共169人169眼。应用配对t检验对两种方法的平均角膜曲率(K)[K=(K1+K2)/2]、前房深度(ACD)分别进行比较,并应用BlandAltman分析对两种测量方法进行一致性评价。结果Pentacam和IOLMaster两种方法的K值分别为(43.35±1.49)D和(43.84±1.54)D,差异具有统计学意义(P<0.01);ACD值分别为(3.65±0.21)mm和(3.70±0.22)mm,差异具有统计学意义(P<0.01)。两种方法的K值一致性较差,而ACD值一致性较好。结论两种仪器在临床应用尤其是进行人工晶体屈光度测量时,角膜曲率值不可相互替代使用,前房深度值可相互替代使用。

关键词: 对比研究, 角膜曲率, 前房深度, IOLMaster, Pentacam

Abstract: To explore the differences of keratometry and antesior chamber depth (ACD) determined by Pentacam and IOLMaster and to evaluate the agreement of the two instruments. Methods169 left eyes of 169 subjects aged (20.80±1.29) with spherical Equivalent-(-3.44±2.04)D were determined keratometry and ACD on both eyes with Pentacam and IOLMaster. Paired t test was used to analyze the difference of mean keratometry (K) [K=(K1+K2)/2] and ACD between the two methods, and the BlandAltman method was used to evaluate the agreement of the two methods. ResultsThe mean K value was(43.35±1.49)D for Pentacam and (43.84±1.54)D for IOLMaster (P<0.01), and the ACD value was(3.65±0.21)mm for Pentacam and(3.70±0.22)mm for IOLMaster(P<0.01). The BlandAltmann showed that the two modalities had incomparable results for K values and comparable results for ACD. ConclusionPentacam and IOLMaster cannot be interchanged for keratometry, but can be interchanged for ACD.

Key words: Pentacam, IOLMaster, Keratometry, Anterior chamber depth, Comparative study

中图分类号: 

  • R778.2