Methodological evaluation of chemiluminescence microparticle
immunoassay for determining HE4

ZHENG Gui-xi, WANG Chuan-xin, LIU Yan-hong, LIANG Yong-yuan   

  1. Department of Clinical Laboratory Medicine, Qilu Hospital of Shandong University, Jinan 250012, China
  • Received:2012-07-06 Online:2012-10-10 Published:2012-10-10

摘要:

目的   对Abbott Architect i2000SR检测系统应用化学发光微粒子免疫技术定量测定人附睾蛋白(HE4)的分析性能进行评价。方法   参考美国临床实验室标准化协会(NCCLS)文件,制定定量检测方法的方法学评价方案,通过Abbott Architect i2000SR检测系统测定HE4的批内精密度、批间精密度、线性范围、携带污染率和参考区间。结果   低、高值混合血清批内精密度CV分别为1.80%、 2.34%;批间精密度CV分别为2.03%、 2.70%;线性范围为19.8~1573.5 pmol/L;携带污染率为-0.02%;参考区间为0.0~140.0pmol/L。结论   Abbott Architect i2000SR检测系统测定HE4的精密度好、线性范围宽、携带污染率低,参考区间与厂家提供的一致,检测性能可满足临床要求。

关键词: Abbott Architect i2000SR检测系统;化学发光微粒子免疫技术;人附睾蛋白4;方法学评价

Abstract:

Objective   To evaluate the performance methodology of Chemiluminescence microparticle immunoassay(CMIA) for determining HE4 using Abbott Architect i2000SR Chemiluminescence analyzer. Methods   According to a method by NCCLS recommended, we designed the evaluation program of quantitative detection methods. The within-run precision、between-run precision、linear range、carry-over and biological reference interval of HE4 were determined. Results   The coefficient of variation(CV) of within-run precision of low and high value pooled serum were respectively 1.80% and 2.34%. The CV of between-run precision were respectively 2.03% and 2.70%. The linear range was from 19.8 pmol/L to 1 573.5 pmol/L. The 1carry-over was -0.02%. The biological reference interval was consistent with manufactures which is from 0.0 pmol/L to 140.0 pmol/L. Conclusion   The withinrun precision, between-run precision, linear range and carry-over of Abbott Architect i2000SR for determining HE4 meet the requirements of clinical use. The biological reference interval was consistent with that provided by the manufacturer.

Key words: Abbott architect i2000SR chemiluminescence analyzer;CMIA;HE4; Methodological evaluation